VHIT NORMS

This material is intended for clinicians and vestibular scientists. See also: bedside version -- HIT test

Timothy C. Hain, MD Page last modified: November 8, 2020

See also: vhit, vhit_conflict, vhit_research, vhit_technical glitches, vhit_compensation.

Norms for the VHIT test -- "all over the place".

One would EXPECT that the VHIT gain in normal people should have a gain 1.0, and there should be few if any saccades during or just after the head thrust. The VHIT gain assumption of 1.0 is wrong because VHIT gains are "all over the place". This is because the VHIT test has a lot of noise.

VHIT gains

The above plot, which we call the "roadkill graph", shows all VHIT tests through 4/2020, from more than 3300 patients tested at Chicago Dizziness and Hearing (using the Interacoustic device). There are many patients with gains > 1. It is almost certain that these values are not "real" but rather are due to measurement error. Thus the VHIT test is not a "clean" test with a normal gain of 1.0, but rather it is "all over the place". Notice that there are also some "real" abnormal VHIT tests shown here -- with low gain on one or both sides. There is useful diagnostic information in VHIT testing, but it is obscured by a cloud of noise, similar to almost all other vestibular tests. This graph was produced using an R-script, shown here.

Why does this graph look like "roadkill" ? Well, there seem to be some "forbidden areas" on this plot -- we just don't have many individuals with gains of 1.0 on one side, and gains < 0.25 on the other. Presumably there are some physiological constrants with superimposed noise. It is likely that this graph shows four groups -- persons who are normal (the big splatter in the middle), people with bilateral vestibular weakness (the tail), and people with unilateral vestibular loss (the wings).

 

Vhit gain Spectacles on VHIT
Figure from Jankey et al (2017) showing scatter in VHIT gains in normal subjects. FIgure from van Dooren et al (2018), showing trend for higher VHIT gain with refractive error.

 

Pursuing this idea, Jankey et al (2017) wrote about "effects of device on .. vHIT gain". They observed that three different VHIT devices all do things a bit differently. They plotted this out in their figure 2, shown above. PG is position gain, IG is instantaneous gain, and AG is area under the curve gain (similar to position gain). Or another way to say this is that position gain is the Visual Eyes device from Micromedical, Instantaeous gain is the EyeSeeCam device from Interacoustic, and Area under curve is the "Impulse" device from GN Otometrics.

This plot shows very clearly that VHIT gains average about 1, but their variability is astounding. About half of normals have VHIT gains greater than 1 ! If this were REALLY TRUE, the eyes would have to go backwards after getting to the target. This doesn't happen often so it isn't true. This means that this variability is NOISE.

So the take home message here is that the 95% limits on VHIT gain looks to be roughly 15% in either direction. We have also observed a similar variability in our clinic population (we use the Interacoustics -- instantaneous gain -- device).

Why are things so variable ? Well it must be that there are lots and lots of technical issues that haven't been worked out. We think ourselves that goggle slippage (i.e. inertia of the camera/google) is the big one. There may also be an effect of the point of regard -- i.e. it has been shown previously using other devices that people who are looking at near targets have higher gains. Finally, perhaps the calibration procedures are not so accurate on these devices. We also wonder whether the video tracking on these devices can be relied upon. Whatever it is, there is a lot of room for improvement.

Rey-Martinez, Burgess and Curthoys (2018) suggested in a case report in an open access journal that "we should consider the hypothesis of a physio-pathologic cause for the enhanced eye responses to vHIT testing of some patients with vestibular function. It is easy to suggest considering hypotheses when one is dealing with one or two cases, but the simplest and most likely assumption, given the wild variability shown above, should just be that the VHIT is a messy test, and that little changes in gain are meaningless.

Van Dooren et al (2018) examined the question as to whether or not use of contacts or spectacles might affect VOR gain. The figure above, from their paper, illustrates their results. In essence, the small effect of refraction was drowned out by the very large variability of VHIT gain. They concluded that it was not necessary to correct for spectacle wearing, arguing that lack of a significant effect was the same as no effect. Of course, the problem with this argument is that had they studied more subjects, it seems highly likely that significance would have been achieved. The difference in gain, from their trend line, looks to be roughly 10% at 5 diopters. Note that these are normal subjects, but their gains are "all over the place".

 


References

  1. Bansal S1, Sinha SK2.Assessment of VOR gain function and its test-retest reliability in normal hearing individuals. Eur Arch Otorhinolaryngol. 2016 Mar 1.
  2. Janky KL, et al. Effects of device on video head impulse test (vHIT) gain. JAAA 28:778-885 (2017).
  3. Rey-Martinez J, Burgess AM, Curthoys IS. Enhanced vestibulo-ocular reflex responses on vHIT.... Frontiers in Neurology (case report), Oct 2018. 1-7 (Note this is both a case report and an open access journal).
  4. Rey-Martinez J, Altuna X, Cheng K, Burgess A, Curthoys IS. Computing endolymph hydrodynamics during head impulse test on normal and hydropic vestibular labyrinth models. Frontieres in Neurology, April 21, 2020. (Note this is an open access journal).
  5. Van Dooren and others. The video head impulse test and the influence of daily use of spectacles to correct a refractive error. Frontiers in Neurology, March 2018. (Note this is an open access journal).