menu Contact Us Dizzy Patients Health Care Providers Research BPPV DVD Tai Chi DVD Understanding Dizziness Acknowledgements Disclaimer Quoting

Rotatory Chair Testing -- Normal Values for sinusoids and step responses

Timothy C. Hain, MD Page last modified: June 3, 2018 button Return to testing index

To discuss what is "normal" for rotatory chairs, we have to first decide on the parameters of interest. There are many possible ways to spin people around, and due to this, many potential "norms".

The "big two" types of rotatory test types are:

System theory says that for a "linear" system, one can represent it as a single "pole", and all of these stimuli should produce the same answers and be interconvertable. It is thought that the vestibular system is linear for low velocities which usually also means low accelerations, but the linearity assumptom is violated at higher velocities. Or in other words, for 30 deg/sec, both steps and sines should give pretty much the same answer. For 300 deg/sec, things could look pretty different due to nonlinearity.

While both the sinusoidal and step responses are very predictable, meaning that people can "suppress" their responses, practically most commercial vestibular testing uses the simpler and less demanding predictable stimuli. By less demanding, we mean that the equipment needed to produce these stimuli can have less torque (power) than that needed to produce predictable stimuli. This tranlates into a more economical device. Imagine for example, that we decided to mix together 5 different sine waves, each one of which that had a maximum acceleration of 100 deg/sec/sec. It would be possible for the sum of these sinewaves to occasionally produce a sine wave that needed 500 deg/sec/sec. Similarly, if each sine had a peak velocity of 60 deg/sec, one could end up with a peak velocity at some point of 300 deg/sec. This is beyond the capability of most commercial rotatory chair devices. Thus it is not important to provide much discussion about SOS or PRBS, as almost nobody uses these protocols.

One might ask then-- why test with both sinusoids and steps, if they are interconvertable ? Basically for redundancy -- when doing something important, it is always a good idea to have two different methods of finding the same answer. I recall in my training, I was talking to Dr. David Robinson (an engineer) about mathematical proofs. Dave told me: Suppose you are building a bridge. Your calculations should also be done two different ways, and you should get the same answer both ways.

Sinusoidal responses:

Figure 1: Typical output of a normal rotatory chair test
normal
Normal rotational chair test. Gain and phase are mainly within normal limits suggested by the manufacturor (Micromedical here).

Normal is defined by gain and phase points staying within the white area on the graph above. Generally speaking, gain points below the lower limit suggest vestibular damage. Gain points above the upper limit suggest a technical error (such as a light leak or uncalibrated chair speed).

So what we have here is a matrix of three outputs from the sinusoidal test, as a function of frequency. The test above then has a total of 7 frequencies*3 values -- 21 numbers. Thats a lot of numbers ! While one can reasonably cut the number of frequencies down to just 4, and also reasonably ignore the asymmetry value, still we have 8 measurements to deal with.

To make things more interesting (?) the frequencies chosen by various authors are not always standardized. In studies that use sum of sines, the frequencies chosen cannot be multiples of each other, as they they would not produce a random stimulus. Instead of going from 0.1 to 0.64, an example might be the study of Peterka et al (1990) where frequencies ranged from 0.0092 to 1.535.

Literature norms for sinusoids:

Ahmed (2014) published a useful set of norms in the Med J Cairo Univ. This article, though not indexed in PubMed, contains the results of testing 100 normal subjects on the Micromedical system 2000 (A commonly used commercial rotatory chair). The age range was 18-56, and there were 66 males and 34 females. Also plotted are the results of Li et al, who studied 41 normal subjects in a similar way, using the Congravez-Goerz model 403D chair. In this study, the higher frequencies were not used (Li et al, 1991).

Note that phase values are almost identical, but gain is noticiably higher for the Li study (done with the Contravez chair), than for the Ahmed study (done with the Micromedical chair).

We do NOT think that setting "norms" for every single frequency is a very useful undertaking, as summary values such as the high-frequency gain and the time constant are far more useful. Still, it is place to start.

Table1/Figure 2: Gain and Phase norms for sinusoidal testing
Freq Gain Phase Calculated Tc Gain Norms Phase norms
.01 .46 37.48 20.75
.02 .57 24.68 17.3
.04 .6 12.37 18.1
.08 .66 4.78 23.8
.16 .63 5.77  
.32 .64 -.91  
.64 .66 -5.99  

Note that the phase values of Ahmed have been multipled by -1 to make them comparable with other studies. The wiggles in these numbers are probably in part artifact or noise. Gain should rise smoothly with frequency. The dip at 0.16 is probably due to noise. Phase should never cross 0, and thus the last value at -5.99 is probably noise as well. As frequency goes up, there is more opportunity for technical error -- the chair might not be able to deliver enough torque, and the goggles might slip on the head.

Commentary about how to convert from Phase to Tc is found below.

Regarding SOS norms

Step responses normal values from literature

Table 2: Gain and TC norms for step testing from various sources
Source Gain Tc n Comment
Ahmed (2014) 0.615

16.39

100 100 deg/sec rotations, averaged starting and stopping
Micromedical 0.535

15.025

  Supplied by manufacturer
Baloh et al (1979) 0.62 18.20   128 deg/sec rotation, data was fit to both Tc and Ta. Fitting for Ta makes Tc larger.
Dizio et al (1990) 0.49 15.48 10 60 deg/sec, Younger individuals in study of age differences
Su et al (2000) 0.53 13.44 55  
Dellepiane et al, 2006 0.42 7.7 20 This is clearly an outlier.
Baloh et al (1993) 0.63 15.1 25 These are the young group from this study of aging

A simple average of the gains and Tc above provides a value of 0.56 and 14.25 for Gain and TC respectively.

 

Tc vs Gain
Figure 3: Literature values for TC/Gains in normal subjects, a bubble chart of data from table 2. The studies with. largest 'n's are the largest bubbles.

 

There is considerable variability here. Possible explanations include:

At any rate, it appears that the two larger studies are the most reasonable, and that the normal Tc should be roughly 15, and gain roughly 0.6

Regarding PRBS norms:

Conversion between Sinusoids and Step Responses

As shown in the table 1 above (Ahmed), one can convert back and forth between the timing numbers (i.e. phase and Tc) for sinusoidal and step responses.

One does need to consider phase sign idiosyncracies. Ordinarily, the eye velocity and head velocity should be equal and opposite at high frequencies (i.e. 0.64). At lower frequencies, the eye velocity is closer to acceleration, and it develops a lead with respect to velocity. While phase leads are usually positive, some authors (like Ahmed above) designate them as negative. You need to figure out whatever convention the paper is using, and if the phases are negative at low frequencies, multiply ALL by -1. Then use this formula:

T=tan(90-phase)/(2pi*F).

If you convert phases in the gain/phase table into time constants, for the lowest 4 frequencies, the numbers are as noted above. Note that the values computed from the sinusoids is not equivalent to the step response time constant, but rather is a little higher. The most reliable is likely the lowest frequency as the phase shift is the largest and thus the most measurable.

The discrepancy in time constants between sinusoids and step responses has been noted by others. There are many possible explanations:

A better way of calculating the time constant is to use ALL of the sinusoidal data, using an iterative curve fit process to a single pole, minimizing the sum of the squares of the error between fit and actual data. This can be easily done in Matlab, using the optimization software, but is beyond the scope of this page. It would be nice if the rotatory chair manufacturers put this into their software.

Gain-TC product -- a method of combining all data into a single number.

We recently published an article regarding using the step response and sinusoids together to quantify vestibular function (Hain, Cherchi and Perez, 2018). This method provides a simple technique to compute the amount of "total" vestibular function. The fraction remaining of total vestibular function is approximately the Gain-TC product/11.3.

Quite a bit more investigation needs to be done to see what GainTc values are in various disease processes.

Effects of Age on Rotatory Chair Norms:

There have been many studies of the VOR in aging, largely concluding that there are modest changes in the VOR gain, TC and phase over age.

Baloh and associates reported the gains and time constants of the VOR in a study of 75 “elderly” normal people, averaging 79.6 years old, who were compared to 25 normal younger people, averaging 26.2 years old (Baloh et al., 1993). The gain, Tc respectively for young was 0.63, 15.1, and 0.58, 11.7 for "Elderly". This documents a small decline with age in both gain and Tc.

Paige et al (1992) reported on 3 age groups, rotated at 0.025 hz, at 50 deg/sec peak velocity.

age (mean) 31 57 79.5
gain 50 0.53 0.49 0.46
phase 50 14.3 17.2 24.6

Clearly, gain gradually declines with age and phase also increases.

Somewhat different than this, according to Maes and associates (2014), children's rotatory chair testing show no effect of age (between 4-12 years of age). In their data, for older persons, there is also surprisingly little effect on rotatory chair testing results, although one would expect a downward trend in gain/phase based on pathological studies showing loss of vestibular hair cells and neurons in the ganglia. There is an effect, but it isn't very large.

Concerning the gain-TC product, which provides a single number to express VOR total response (Hain et al, 2018)

Baloh and associates reported the gains and time constants of the VOR in a study of 75 “elderly” normal people, averaging 79.6 years old, who were compared to 25 normal younger people, averaging 26.2 years old (Baloh et al., 1993). The gain:Tc for young was 0.63:15.1, and 0.58:11.7 for "Elderly". Their data results in a GainTc product of 9.5 for the younger subjects, and 6.8 for the older subjects, from which it can be calculated that the slope of the GainTc/year is -0.051/year.

Similarly, Paige (Paige, 1992) reported gain and phase data in 30 “young” (18-44) and 23 “Middle-Aged” (45-69) subjects. The GainTc products for the 0.025Hz -- 50 deg/sec stimulus, similar to our methodology, were computed to be 13.23 and 10.07 for young and middle-aged, respectively, resulting in a slope of -0.12 deg/year. This analysis suggests that the GainTc product declines slowly with age.

References:

Copyright June 3, 2018 , Timothy C. Hain, M.D. All rights reserved. Last saved on June 3, 2018